The CDD Linear X-ray detector is an integrated detection module with a linear diode array up to 1024mm in length.
It has the following features :
- Available in diode pitch of 1.6mm, 0.8mm, 0.4mm, 0.2mm
- Data interface RS422 pixel rate of up to 4MHz
- Scan rate from 50 to 4000 scans/sec
- Output resolution of 8 to 12 bits as standard
- Control interface USB 2.0 as standard, RS232 optionally
- Multi channel radiographic trigger
- Scan rate fixed or synchronous to external clock
- Pixel output normalised by internal DSP processor
- Output translation tables provide gamma law
- All cables exit from end of module for easy connection
Control Commands
Set Line/Integration period
Zero diodes (set black level)
Full range for diodes (set maximum white level)
Set gamma law lookup table
Start/stop scanning
Display raw (input) data
Display normalised output data
Display underlying zero, gain and multiplier data
Histogram display of pixel grey levels
Calculate noise for each pixel
Display test pattern output
Radiographic trigger setup
General Specifications:
| |
CDD XX2 |
CDD XX4 |
CDD XX8 |
CDD XX16 |
| X-ray tube kVp range |
20-160 kVp |
20-160 kVp |
20-160 kVp |
20-160 kVp |
| Scintillator |
GOS |
GOS |
GOS |
GOS |
| Active area lengths |
102-410 mm |
102-819 mm |
102-1024 mm |
205-1638 mm |
| Number of pixels |
512-2048 |
256-2048 |
128-1280 |
128-1024 |
| Pixel Pitch |
0.2 mm |
0.4 mm |
0.8 mm |
1.6 mm |
| Pixel Height |
0.3 mm |
0.6 mm |
0.8 mm |
1.6 mm |
| Pixel Width |
0.1 mm |
0.3 mm |
0.7 mm |
1.4 mm |
| Max. Scanning Speed |
12 m/mn (1) |
50 m/mn (1) |
170m/mn (1) |
300m/mn (1) |
| Min. Integration Time |
1 ms |
0.5 ms |
0.25 ms |
0.25 ms |
| Max Integration Time |
20 ms |
20 ms |
20 ms |
20 ms |
| A/D Resolution |
12-bits |
12-bits |
12-bits |
12-bits |
| Dynamic Range |
> 2000 (2) |
> 2000 (2) |
> 2000 (2) |
> 2000 (2) |
| Data Digital Interface |
12-bits (3) |
12-bits (3) |
12-bits (3) |
12-bits (3) |
| Power consumption |
50 W max |
50 W max |
50 W max |
50 W max |
| Image acquisition interface |
RS-422 |
RS-422 |
RS-422 |
RS-422 |
| Control interface |
USB 2.0 |
USB 2.0 |
USB 2.0 |
USB 2.0 |
Notes
- Maximum speed is specified for active length up to 512 mm.
- Dynamic Range is defined as saturation signal/RMS noise, normal detector sensitivity.
- Digital interface has 16 RS-422 I/O lines for data, but user can select actual data to be 14,12,10 or 8 bits. The default value is 8 or 12 bits
- User can designate radiographic trigger outputs by specifying Window left, Window right and Threshold.
Physical Dimensions:
| Model |
Active Length |
Number of Pixels |
Length |
Width |
Height |
Weight |
| CDD048 |
204mm |
256 |
260mm |
130mm |
50mm |
4Kg |
| CDD068 |
307mm |
384 |
360mm |
130mm |
50mm |
6Kg |
| CDD098 |
461mm |
576 |
510mm |
130mm |
50mm |
9Kg |
| CDD128 |
614mm |
768 |
660mm |
130mm |
50mm |
12Kg |
Enviromental Characteristics:
| Characteristic |
Specific characteristic |
Specification |
Notes |
| Operational |
Ambient temperature |
0-40 C |
|
| |
Relative humidity |
30-80 % (non condensing) |
|
| Storage |
Ambient temperature |
-10-50 C |
|
| |
Relative humidity |
30-80 % (non condensing) |
|
Power Requirements:
| Supply voltage |
Maximum current |
| + 5 volts |
5.0 Amp |
| +12 volts |
1.0 Amp |
| - 12 volts |
0.2 Amp |
Ordering Information:
Options:
-
R RS232 control interface
-
Bnn Output resolution (bits)